Abstract

This paper demonstrates that glucose determination in blood can be done directly (without sample pretreatment) using a reagentless reversible biosensor based on the intrinsic spectroscopic properties of peroxidase (HRP). The biosensor, prepared by HRP and glucose oxidase entrapment in a polyacrylamide gel matrix, works in continuous mode, presents a linear response range from 1.5×10−6 up to 5.5×10−5M and can be used for at least 750 measurements; in the best conditions (0.1M pH 6 phosphate buffer, HRP and GOx amounts in the polymersation mixture for the sensor film preparation 0.0165 and 0.0010g, respectively) the minimum samples rate is 30h−1. For glucose determination, blood is simply diluted in water (until haemolysis is completed) and fed into the sensor without a cleaning step between samples; the blood absorption is corrected in a simple way by working at a proper reference wavelength. The biosensor signals have been mathematically modeled in order to facilitate the design of sensors based on the same idea for other biochemical compounds.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.