Abstract

A major obstacle for characterizing dissolved organic matter (DOM) with ultrahigh resolution mass spectrometry has been its low concentrations in natural waters. Many previous mass spectrometric studies of both terrestrial and marine DOM typically have isolated and concentrated the DOM using solid phase C18 extraction disks, ultrafiltration, or XAD resins, all of which are known to discriminate against many different classes of compounds. We have, for the first time, developed an approach to directly analyze natural water samples with less than 6mg/l DOC (dissolved organic carbon), using electrospray ionization Fourier transform ion cyclotron resonance mass spectrometry (ESI-FTICR-MS). We demonstrate the sensitivity and ability of sequential selective ion accumulation (SSIA) to detect the thousands of components in a single freshwater DOM sample without any significant pretreatment. By utilizing SSIA, the baseline noise decreases while signal to noise ratios of the peaks increase, allowing for approximately 40% more formulas to be assigned to peaks in the mass spectra.