A novel reagentless direct electrochemical DNA sensor has been developed using ultrathin films of the conducting polymer polypyrrole doped with an oligonucleotide probe. Our goal was to develop a prototype electrochemical DNA sensor for detection of a biowarfare pathogen, variola major virus. The sensor has been optimized for higher specificity and sensitivity. It was possible to detect 1.6 fmol of complementary oligonucleotide target in 0.1 ml in seconds by using chronoamperometry. The sensitivity of the developed sensor is comparable to indirect electrochemical DNA sensors, which use electrochemical labels and reagent-intensive amplification. The developed sensing electrode is reusable, highly stable and suitable for storage in solution or in dry state.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.