Abstract

We report linear sweep and square wave voltammetric studies on glucose oxidation at boron-doped diamond (BDD) electrodes in an alkaline medium in efforts to evaluate the techniques for electrochemically assaying glucose. The bare BDD electrode showed good linear responses to glucose oxidation for a concentration range from 0.5 to 10 mM glucose, which well encompasses the physiological range of 3–8 mM. The BDD electrodes did not experience interferences from ascorbic acid or uric acid during glucose detection. This method, when applied to real blood samples, gave results similar to those obtained by a commercial glucose monitor.