Abstract

We describe the development of a direct method for determination of selenium in serum by graphite-furnace atomic absorption spectrophotometry with deuterium background correction. We include palladium modifier to stabilize selenium in the presence of a strong reducing agent. Spectral interferences from iron are not evident in this system. Because an analysis requires only 20 microL of serum or plasma, the method is applicable to neonatal populations. At a selenium concentration of 0.75 mumol/L (approximately the mean found in our pre-term infants), the within-run CV is 5.3%. For a concentration of 1.83 mumol/L, which approximates our normal adult mean, within-run and between-run CVs are 2.7% and 3.4%. Accuracy is demonstrated by analytical recoveries ranging from 96% to 102%. We present reference values for pre-term and term infants, and age-specific ranges for infants to adults.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.