Abstract

A procedure is proposed for the direct determination of manganese and iron in wine employing fast sequential flame atomic absorption spectrometry and the reference element technique to correct for matrix effects. Cobalt, silver, nickel and indium have been tested as reference elements. The results demonstrated that cobalt and indium at a concentration of 2 and 10 mg L −1 were efficient for quantification of manganese and iron, respectively. Under these conditions, manganese and iron could be determined with quantification limits of 27 and 40 μg L −1, respectively. The proposed method was applied to the determination of manganese and iron in 16 wine samples. The content of manganese varied from 0.78 to 2.89 mg L −1 and that of iron from 0.88 to 9.22 mg L −1. The analytical results were compared with those obtained by inductively coupled plasma optical emission spectrometry after complete mineralization using acid digestion. The statistical comparison by a t-test (95% confidence level) showed no significant difference between the results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.