Abstract

The 1-octyl-3-methylimidazolium hexafluorophosphate ([C 8MIM][PF 6]) ionic liquid was immobilized in the pores of a polypropylene hollow fiber for hollow fiber-protected liquid-phase microextraction. Analytes including 4-chlorophenol (4-CP), 3-chorophenol (3-CP), 2,4-dichlorophenol (2,4-DCP) and 2,4,6-trichlorophenol (2,4,6-TCP) were extracted into this ionic liquid membrane, and back extracted into 10 μL sodium hydroxide acceptor solution in the lumen of the hollow fiber. Then, the acceptor solution was withdrawn into the high-performance liquid chromatography (HPLC) microsyringe connected to the hollow fiber, and directly injected into the HPLC system for analysis. Some parameters that might affect the extraction efficiency were optimized, and low detection limits (0.5 μg L −1 for 4-CP, 3-CP, DCP and 1.0 μg L −1 for TCP) were obtained. Good repeatability was achieved because of the stability of the hollow fiber-supported ionic liquid membrane. The proposed procedure was applied for direct determination of the four chlorophenols in some real water samples including groundwater, river water, wastewater and tap water. All of the four chlorophenols in these water samples were under the limits of determination, and the recoveries were in the range of 70.0–95.7% at 5 μg L −1 spiked level.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call