Abstract
A procedure for the determination of cadmium in fresh, coastal and estuarine waters by polarized Zeeman-effect graphite-furnace atomic absorption spectrometry is validated by using lake waters and seawater. The limit of detection for freshwaters is <2 ng l −1 cadmium. Undiluted seawater can be analyzed directly without the addition of matrix modifiers with the aid of a stabilized temperature platform. The instrument is calibrated with diluted NBS SRM 1643a (Trace Elements in Water). Analytical performance was tested extensively with fresh and brackish water samples and procedures were worked out to ensure that a high degree of accurately is achieved consistently.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.