Abstract
Pump induced hemolysis is presently evaluated by measuring plasma free hemoglobin (fHb). However, this method has disadvantages because quantification of fHb depends on hematocrit (HCT) and hemoglobin (Hb) levels. The aim of this work was to devise a hemoglobin independent method, capable of quantifying cell trauma directly by measuring the number of red blood cell (RBC) fragments. Whole blood flow cytometry was used to quantify circulating RBC fragments derived from a roller pump (Sarns, Inc. Model 2 M 6,002) and a centrifugal pump (Gyro C1E3, Kyocera Corp.). The pumps were tested in a mock circuit for 2 hr (5 L/min flow against 100 mm Hg pressure head). Red blood cell fragments were quantified by a phycoerythrin (PE) labeled glycophorin A antibody specific for erythrocytes. Red blood cell fragments were smaller than the intact RBC population and overlapped in size with the platelet population (based on forward- and side-light scattering measurements). For the roller pump, the values for RBC fragments increased from 1,090 +/- 260/microl at 0 min to 14,880 +/- 5,900/microl after 120 min. In contrast, using the centrifugal pump, there was little increase in RBC fragments (from 730 +/- 270/microl at 0 min to 1,400 +/- 840/microl after 120 min). Flow cytometry can be used for the rapid, sensitive, hemoglobin independent evaluation of pump induced RBC trauma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: ASAIO journal (American Society for Artificial Internal Organs : 1992)
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.