Abstract

A simple system for stabilizing a scintillation counter is described which uses a dc light source (a green light emitting diode) to illuminate the photo-cathode of the photomultiplier used to detect γ-induced light pulses from the scintillator. Basically, the photomultiplier anode current due to the light emitting diode light is held constant by an automatic control loop acting on the eht voltage to keep the gain of the photomultiplier tube constant. However, because the temperature coefficient of the scintillator does not in general match that of the light emitting diode, further compensation is required to achieve constant γ pulse gain. This is provided by adding to the control circuit a current derived from the light emitting diode voltage which is an excellent measure of temperature; the use of this technique results in gain constancy to within ±1% in the 10–50°C ambient temperature range. Noise and countrate limitations are discussed and it is concluded that the system is generally applicable to uranium prospecting equipment.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.