To assess the outcomes of direct coronary stenting (DS) using long stents and examine predictive factors of DS failure, this prospective multicenter registry included 128 consecutive patients who underwent the implantation of stents >or= 18 mm in length without balloon predilation for de novo coronary artery stenoses. Mean lesion and stent lengths were 20.7 +/- 5.4 and 21.4 +/- 3.8 mm, respectively. Rates of DS success, lesion success, and primary success were 82%, 99%, and 97.7%, respectively. At 6 months, rates of MACE and TVR were 12.5% and 6.3%, respectively. In multivariate analysis, factors predictive of DS failure vs. DS success were presence of calcifications (78% vs. 45%; P = 0.004) and reference vessel diameter (2.77 +/- 0.4 vs. 3.13 +/- 0.42 mm; P = 0.0002). DS of complex lesions with stents >or= 18 mm in length was performed safely and with a high success rate. This strategy was less successful in the treatment of small vessels and in presence of calcifications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.