Abstract

BackgroundRapid molecular diagnostics by PCR has a crucial role in handling the global SARS-CoV-2 pandemic. As diagnoses are time-sensitive and global supply chains are susceptible to various factors alternative detection methods would be an important backup. ObjectivesDuring the study the performance of a commercially available isothermal LAMP method for SARS-CoV-2 detection was compared to a IVD RT-PCR Assays using throat wash specimens that were routinely taken in our hospital setting. Study designThroat wash specimens of hospital staff (n = 174) previously tested positive for SARS-CoV-2 by the Altona Diagnostics RealStar SARS-CoV-2 RT-PCR (Altona Diagnostics, Hamburg, Germany) was tested for SARS-CoV-2 also by the SARS-CoV-2 Rapid Colorimetric LAMP Assay (NEB Germany GmbH, Frankfurt a.M., Germany). ResultsThe sensitivity of the colorimetric LAMP Assay compared to RT-qPCR was 78.74%, and the specificity was determined to 88.24% with a positive predictive value of 0.986 and a negative predicitve value of 0.882. The positive and negative likelihood ratio for LAMP was 6.693 and 0.241, respectively, while the diagnostic odds ratio was 27.77. ConclusionsIn times of limited PCR test ressources and in settings with limited PCR capacities, the colorimetric LAMP Assay could serve as an alternative, if a calculable loss of sensitivity is acceptable from the Public Health perspective in certain settings.