Abstract

PurposeSurface imaging (SI) offers a nonionizing, near real time alternative to radiographic imaging for intrafraction radiosurgery localization. In this work, we systematically compared a commercial SI system vs a commercial room mounted x‐ray localization system in phantom.MethodsAn anthropomorphic head phantom with fiducial markers was imaged with linear accelerator on‐board x‐ray imaging, SI, and room mounted x‐ray imaging (RM) at ±45° and ±90° couch angles for three different head tilts and six different isocenters (72 total positions). The shifts generated by the three systems were compared as functions of couch angle, head tilt, and isocenter position with the on‐board imaging shifts used as ground truth. Two sample Kolmogorov–Smirnov tests were used to evaluate equivalence of the groups.ResultsThe magnitude of the displacement vectors for RM minus on‐board imaging and SI minus on‐board imaging over all 72 phantom positions were 0.7 ± 0.3 mm for both cases. The RM and SI showed no significant difference based on couch angle or isocenter position. Both systems showed decreasing accuracy with increasing couch angle, but both systems agreed with ground truth to <=1.1 mm at all couch angles. The exaggerated chin‐up head orientation showed significantly different shifts for SI and RM based on increased variance in the SI measurements, although both had submillimeter accuracy on average. The standard deviation of the real time SI displacement vector was <0.06 mm over all measurements, during which the on‐board imaging panels partially blocked the lateral camera pods for half the time.Conclusions RM and SI showed similar accuracy over measurements at 72 different phantom positions. SI showed minimal performance loss with camera pods blocked. SI is a feasible option for intra‐fraction radiosurgery localization based on these phantom measurements.

Highlights

  • Stereotactic radiosurgery (SRS) has become a popular tool to treat intracranial brain metastases due to durable local control, convenience for the patients, and the possibility of reduced cognitive impairment vs whole brain radiotherapy.[1,2] One of the main technical recommendations for SRS is that patients can be localized to

  • We found that ART and ET showed comparable positional accuracy at all couch angles, and that both systems showed decreasing accuracy with increasing couch angle

  • We found a mean ART SD of 0.06 mm, which included measurements with both TrueBeam STx (TB) imaging arms extended partially blocking the lateral cameras

Read more

Summary

Introduction

Stereotactic radiosurgery (SRS) has become a popular tool to treat intracranial brain metastases due to durable local control, convenience for the patients, and the possibility of reduced cognitive impairment vs whole brain radiotherapy.[1,2] One of the main technical recommendations for SRS is that patients can be localized to

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.