Abstract
Abstract A simple and sensitive chemiluminescence (CL) method was developed for direct quantification of oxymorphone, a µ opioid agonist that is approximately 10 times more potent than morphine. In this method potassium permanganate in polyphosphoric acid was used as CL reagent. Using this method, oxymorphone can be determined over the concentration ranges 13.5-337.8 ng mL-1 and 0.34-6.76 µg mL-1 with a sampling rate of 45 samples h−1. The limit of detection was 3.5 ng mL-1 (signal to noise =3) and the percent of relative standard deviations (in 9 replicate measurements) were 3.1% for 135.1 ng mL-1 and 3.6% for 1.4 µg mL-1 oxymorphone. The method is applied to human plasma and synthetic samples. The CL mechanism has been proposed using UV-Vis, fluorescence and CL spectra. In this study, detectability of oxymorphone in some other CL systems such as Ce(IV)-H2SO4, luminol-H2O2, Ru(phen)32+-Ce(IV) and permanganate-SO32- is investigated. CL intensities of twelve narcotics or related drugs were also investigated in the proposed CL system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.