Abstract

A totally implantable pulsed Doppler ultrasonic blood flowmeter has recently been developed to provide information on the velocity-flow profile in a vessel as well as its diameter. Volume flow can be indirectly obtained according to the formula: Q = (phi/4)(diam)2-v. In order to determine the accuracy of this estimate, in vivo direct bleedout measurements were performed on the abdominal aorta of six dogs with an overall accuracy in 77 trials of +2.0 +/- 8.7% (mean +/- 1 SD). The best-fit mean linear regression line was found to be: flowmeter output = 1.013-true flow + 5.1 ml/min. The scatter in the flowmeter's accuracy is thought to be due to small variations in the angle of the transducer. The source of the 2% overestimation in the mean accuracy could not be directly attributed to any one condition; the error is sufficiently small that in many cases it can be neglected.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.