Abstract

Atmospheric pressure chemical ionization-tandem mass spectrometry (APCI-MS/MS) was applied for the first time to the direct analysis of octamethylcyclotetrasiloxane (D4) and decamethylcyclopentasiloxane (D5) in gaseous matrixes without extraction or prior chromatographic separation. Mass spectrometric characteristics of both compounds under APCI conditions and their fragmentation behavior in MS/MS were investigated. Unlike the classical gas chromatography/mass spectrometry (GC/MS), which involves solvent extraction before gas chromatography, the proposed approach prevents contamination from GC system components and provides unambiguous structural assignments. The method performs well achieving good linearity (R(2) > 0.997), low limits of detection (4-6 microg/m(3)), good precision (RSD < 10%) and accuracy (>93%), and a wide dynamic range. Its applicability to real-world samples was evaluated through measurements of D4 and D5 concentrations in air and biogas samples. The high sensitivity, selectivity, and reliability of this method render our approach a good alternative to the commonly used GC/MS method.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.