Abstract

A method is described for the direct measurement of Kryptofix 2.2.2 levels in 18FDG preparations dissolved in isotonic saline using gas Chromatography coupled with a nitrogen-selective detector. The cryptand can be eluted intact in 2 min time from a short megabore capillary column, and detected at levels as low as 0.25 μg/mL. This method offers rapid, quantitative and sensitive detection of Kryptofix 2.2.2 levels in 18FDG thus making it amenable to routine quality assurance testing.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.