Abstract

In 2007, the United States Food and Drug Administration released guidance recommending testing of glycerin used in regulated consumer products, such as cough syrup preparations, toothpaste, and other pharmaceutical and food products, for the toxic compounds ethylene glycol and diethylene glycol. Regulatory laboratories routinely test glycerin, and products containing glycerin or related compounds for these toxic glycols, using an official gas chromatographic method, to ensure the safety of these products. The current work describes a companion technique to compliment this GC-FID method utilizing Orbitrap mass spectrometry with direct analysis in real time ionization to rapidly screen these samples qualitatively, with results in as little as five seconds, with no sample preparation required. This allows the more time and resource intensive method to be reserved for those rare cases when these compounds are detected, potentially greatly improving laboratory efficiency. The technique was evaluated for qualitative sensitivity and repeatability, and compared against the GC-FID method. The method appears to perform well against these metrics.