Abstract
It was determined that employment of the Newark Beth Israel Hospital (NBIH) intracardiac dipolar electrode for temporary cardiac pacing in preparation of patients for permanent pacemaker implantation or other operation is a dependable and easily performed procedure. The dipolar electrode was found most effective when positioned at any site in the right ventricle where the lowest driving voltage is required. Endocardial contact was not necessary. The optimal distance between electrodes was 1 cm or less, the need thus being obviated for an external indifferent electrode. Endocardial or myocardial damage was not demonstrated. Subsequent major operation was uneventful when cardiac rate and rhythmicity were controlled. Confidence of the patient was markedly improved. Poor-risk patients were well supported. Patients were ambulatory. Formation of clot on the positive pole of a tripolar or coaxial catheter was not demonstrated in animals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.