Abstract

Recently, space radiation dosimetry measurements were made by passive and active detectors inside the Spacelab [STS-47 (FMPT): 300km, 57°, STS-65 (IML-2 mission): 300km, 28.5°]. The LET distributions obtained by antioxidant doped CR-39 inside the Spacelab were compared with those measured by the tissue equivalent proportional counter (TEPC) and the real time radiation monitoring device (RRMD) consisting of eight silicon detectors. While both distributions by CR-39 are in good agreement with those obtained by active detectors over the region of LET of several tens to 200 keV/μm, a significant difference in the LET region of smaller than several tens keV/μm is seen. It is considered to be caused by the dip angle dependence of track formation sensitivity in CR-39. The track formation sensitivity for different dip angle were measured for several high heavy energy ions. Using these results, the correction for the dip angle was made for the LET distribution. The corrected result is consistent with the results obtained by active detectors.