Abstract

Microplate reader assays offer several advantages over conventional spectrophotometric assays. We adapted the dihydrofolate reductase (DHFR) enzyme inhibition assay for use in a 96-well microplate reader to measure plasma methotrexate (MTX) concentrations. The assay is linear from 0.01 to 0.1 micromol/L. The within-run CVs at 0.03 micromol/L and 0.08 micromol/L MTX were 4.0% and 2.7%, respectively, and the interday (total) CVs were 7.6% and 1.8%. Cross-reactivity with the inactive MTX metabolite 2, 4-diamino-N10-methylpteroic acid (DAMPA) was 3.9%, significantly less than that described with commercial immunoassays; with 7-hydroxymethotrexate cross-reactivity was 1.7%. In addition to sensitivity and specificity, the advantages of this assay are small sample volumes, simultaneous analysis of multiple samples, and rapid turnaround. Because of its greater specificity, the DHFR enzyme inhibition assay may be useful when DAMPA is present in plasma samples and HPLC is not available.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.