Abstract
The aim of this study of native pig mandibles was to investigate the accuracy and quality of the representation of peri-implant defects by intraoral radiography (IR), panoramic radiography (PR), computer tomography (CT), and digital volume tomography (DVT). The examination was carried out on 19 native pig mandibles. In the toothless sections of the mandibles, one or two implants were inserted. Following the standardized preparation of peri-implant defects (11 each of dehiscences, fenestrations, and 2- to 3-walled intrabony defects), IR, PR, CT, and DVT were performed. The peri-implant defects were measured using appropriate software on the digitized IR and PR image programs. As a control method, the peri-implant bone defects were measured directly using a reflecting stereomicroscope with measuring ocular. The statistical comparison between the measurements of the radiographic scans and those of the direct readings of the peri-implant defects was performed with Pearson's correlation coefficient. The quality of the radiographic scans was determined through the subjective perception and detectability of the peri-implant defects by five independent observers. In the DVT and CT scans, it was possible to measure all the bone defects in three planes. Comparison with the direct peri-implant defect measurements yielded a mean deviation of 0.17+/-0.11 mm for the DVT scans and 0.18+/-0.12 mm for the CT scans. On the IR and PR images, the defects could be detected only in the mesio-distal and cranio-caudal planes. In comparison with the direct measurements of the peri-implant defects, the IR images revealed a mean deviation of 0.34+/-0.30 mm, and the PR images revealed a mean deviation of 0.41+/-0.35 mm. The quality rating of the radiographic images was highest for the DVT scans. Overall, the CT and DVT scans displayed only a slight deviation in the extent of the peri-implant defects. Both radiographic imaging techniques permitted imaging of peri-implant defects in three planes, true to scale, and without overlay or distortion. The DVT scans showed the best imaging quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.