Abstract

A videodensitometric method for quantification of aortic regurgitation which requires neither measurement of cardiac output nor determination of enddiastolic and endsystolic left ventricular volumes has been developed. The injection of 20 ml of contrast medium into the left ventricle is digitally recorded at 25 images s-1 during 20 s using an equipment for digital subtraction angiography (Digitron 2, Siemens). The Digitron computes 2 'time dilution curves' (TDC) from the unsubtracted image sequence, for 2 regions of interest drawn around the angiographic enddiastolic and endsystolic left ventricular silhouettes. Enddiastolic and endsystolic points of the TDC are then entered into a VAX-750 computer, which calculates the ejection fraction (EF), the forward ejection fraction (FEF) and the regurgitant fraction (RGF). This is performed by a complex fitting algorithm based on a physical model of the washout process of contrast medium, which reconstructs the two best enddiastolic and endsystolic baselines in the washout parts of the two TDC. The EF, FEF and RGF obtained in 9 regurgitant and 11 nonregurgitant patients have been compared with the corresponding values EFv, FEFv and RGFv obtained by a conventional technique (Cardiogreen and biplane LV area-length volumetry). Regression analysis yielded: EF = 0.88 X EFv (regression line forced through the origin), r = 0.77, FEF = 0.76 X FEFv + 3, r = 0.96, RGF = 0.94 X RGFv + 5, r = 0.98 (v stands for volumetry).