Abstract
The airflow balance test to confirm the physical performance of biological safety cabinets (BSC) is standardized by JIS K 3800 in accordance with NSF/ANSI 49 in Japan. In the test, Bacillus atrophaeus spores are used as an evaluation index, which poses a few limitations, such as their dependence on the environmental conditions to be tested. The aim of this study is to develop BSC digital twins and discuss reasonable alternatives for performance evaluation tests while maintaining the accuracy of the bacterial test performance. Herein, we provide the outline of a digital twin model for BSCs and the results of a numerical analysis, where a passive scalar is used as an alternative pollutant in the evaluation index.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.