Abstract

Monitoring the stability of patient position is essential during high-precision radiotherapy such as spine stereotactic body radiotherapy (SBRT). We evaluated the combination of digital tomosynthesis (DTS) and triangulation for spine position detection, using non-clinical DTS software and an anthropomorphic pelvic phantom that includes a bone-like spine structure. Kilovoltage cone beam CT projection images over 2–16° gantry rotation were used to generate single slice DTS images. Each DTS slice was registered to a digitally reconstructed DTS derived from the planning CT scan to determine 2D shifts between actual phantom and treatment plan position. Two or more DTS registrations, central axes 4–22° apart, were triangulated to determine the 3D phantom position. Using sequentially generated DTS images, the phantom position can be updated every degree with a small latency of DTS and triangulation angle. The precision of position determination was investigated as function of DTS and triangulation angle. To mimic the scenario of spine SBRT, the effect on the standard deviation of megavoltage radiation delivery during kV image acquisition was tested. In addition, the ability of the system to detect different types of movement was investigated for a variety of small sudden and gradual movements during kV image acquisition.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.