Abstract

The value of digital radiography in musculoskeletal disorders was investigated by assessing its ability to depict anatomic structures and common radiologic features as compared with the conventional film-screen combination. the digital image that was frequency modified was superior to conventional films in delineating soft tissue structures and for areas with large attenuation differences. the conventional film-screen system was superior in depicting small anatomic structures and in identifying the zone close to prostheses. This was explained by the high spatial resolution of the conventional film system and the disturbing halo effect around the prosthesis seen with digital images. the halo effect is an overshoot caused by the unsharp masking operator, which was in this series not changed for individual examinations. the exposure (radiation dose) could be reduced to 50 per cent using the digital system, without any loss of information

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.