Abstract

A technique is described for lateral pelvimetry which results in a lower radiation dose than that resulting from conventional film-screen pelvimetry and computed tomography (CT) pelvimetry. It uses digital radiographic imaging thereby closely matching conventional techniques yet additionally providing post-processing facilities including distance measurement and contrast adjustment. The digital technique was evaluated with an anthropomorphic phantom and compared with the conventional technique by measuring entrance surface doses on two groups of post-natal patients. Digital radiography gave radiation doses 10 times lower than those using conventional techniques.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.