We have developed a real-time digital pulse-shape analysis system for a Ge Compton camera. We successfully demonstrated the improvement of the intrinsic position resolution using the digital pulse-shape analysis. A test measurement with a 54Mn point source showed that the spatial resolution of the Compton camera with the digital system were improved about 1.5 times. The first imaging of an RI phantom, that has 2D hot spots filled with 64Cu of 3 MBq, demonstrated that few millimeter spatial resolution was achieved with our new system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.