Abstract

In this research, based on pulse reconstruction method, digital storage oscilloscopes with an 8-bit digital-to-analog convertor was used to successfully perform the neutron-gamma discrimination with NE213 (or its equivalent BC501A) scintillator anode pulses at minimum discrimination bias value of 95 keVee (or keV electron equivalent). Also, a 100 mCi 241Am–Be source and micro-Curie gamma ray sources (137Cs and 22Na) were used for the system calibration and discrimination studies. The results confirm the feasibility and simplicity of the proposed method.