Abstract

In this research, based on pulse reconstruction method, digital storage oscilloscopes with an 8-bit digital-to-analog convertor was used to successfully perform the neutron-gamma discrimination with NE213 (or its equivalent BC501A) scintillator anode pulses at minimum discrimination bias value of 95 keVee (or keV electron equivalent). Also, a 100 mCi 241Am–Be source and micro-Curie gamma ray sources (137Cs and 22Na) were used for the system calibration and discrimination studies. The results confirm the feasibility and simplicity of the proposed method.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.