Abstract

The accuracy of digital luminescence radiography was compared with that of conventional film/screen techniques, using animal preparations and clinical examinations. Fine fissures were made in 8 animal bones and these were examined radiologically. The digital examinations were carried out with and without edge enhancement. 208 patients were examined in a similar way. Film quality and assessment of the fractures were evaluated quantitatively. In no instance did either of the digital methods provide inferior quality when compared with conventional films. ROC analysis for evaluation of fractures in patients, using an experienced radiologist, showed no significant difference between the various methods (ROC areas: conventional 0.947, digital 0.958, digital with edge enhancement 0.943). With a less experienced observer there were significant advantages for both digital methods (ROC areas: 0.851, 0.886, 0.908). Our investigation has proved that fractures which are difficult to see can be reliably demonstrated by digital luminescence radiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.