Abstract

The aim of this study was to compare the diagnostic quality of conventional cephalometric radiographs with that of digital image counterparts. The random error associated with angular and linear measurements recorded on the digital images was greater than on the conventional radiographs. In addition, there was a systematic error producing statistically significant differences in the majority of angular and linear measurements between the digital images and the conventional radiographs. The errors that occurred with some measurements were of sufficient magnitude to be of clinical significance, particularly in a cephalometric situation where a high degree of accuracy is required. It is therefore suggested that, for digital imaging of cephalometric radiographs, a pixel matrix larger than 512 x 512 with more than 64 gray levels is required to maintain the diagnostic quality of the original radiograph.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.