Abstract
Digital Image-based Elasto-Tomography (DIET) is a novel surface-based elasticity reconstruction method for determining the elastic property distribution within the breast. Following on from proof of concept simulation studies, this research considers the motion evaluation and stiffness reconstruction of a soft tissue approximating gelatine phantom. This initial phantom work provides an intermediate stage between prior simulation studies more detailed phantom studies to follow. Reference points on the surface of a cylindrical phantom were successfully tracked and converted into a steady-state motion description. Motion error based mechanical property reconstruction allowed an estimation of the stiffness of the gelatine when actuated at 50Hz. The reconstructed stiffness compared favorably with independently measured stiffness properties of the gelatine material when experimental assumptions were considered. An experimental noise estimate of 50% was confirmed accurate by comparing experimental motions to simulated motion data with added noise.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.