Abstract

Digital image analysis (DIA) is used increasingly as an assisting tool to evaluate biomarkers, including human epidermal growth factor receptor 2 (HER2) in invasive breast cancer (BC). DIA can assist pathologists in HER2 evaluation by presenting quantitative information about the HER2 staining in APP assisted reading (AR). Concurrently, the HER2-low category (HER2-1+/2+ without HER2 gene amplification) has gained prominence due to newly developed antibody-drug conjugates. However, major inter- and intraobserver variability have been observed for the entity. The present quality assurance study investigated the concordance between DIA and AR in clinical use, especially concerning the HER2-low category. HER2 immunohistochemistry (IHC) in 761 tumours from 727 patients was evaluated in tissue microarray (TMA) cores by DIA (Visiopharm HER2-CONNECT) and AR. Overall concordance between HER2-scores were 73% (n=552, weighted-κ: 0.66), and 88% (n=669, weighted-κ: 0.70), when combining HER2-0/1+. A total of 205 scores were discordant by one category, while four were discordant by two categories. A heterogeneous HER2 pattern was relatively common in the discordant cases and a pitfall in the categorisation of HER2-low BC. AR more commonly reassigned a lower HER2 score (from HER2-1+ to HER2-0) within the HER2-low subgroup (n=624) compared with DIA. DIA and AR display moderate agreement with heterogeneous and aberrant staining, representing a source of discordance and a pitfall in the evaluation of HER2.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call