Abstract

Images obtained from an analog SEM are traditionally viewed and recorded from a cathode-ray tube (CRT). Many laboratories use instant film (e.g. Polaroid #52, #55 instant film) to justify image quality and obtain permanent image quickly. Digital imaging provides an alternative approach for image acquisition and recording. One major advantage of digital SEM is image averaging that allows one to improve the signal-to-noise ratio (SNR) from a noisy quick-scan image to reduce charging. SEM signal yield is proportional to incident beam intensity, image acquisition time or duration of beam interaction with specimen (dwell time). The higher beam intensity, or longer the dwell time, the more signal generated. However, for high-resolution SEM imaging, the beam dose and dwell time are limited by drafting, radiation damage, and contamination. Therefore high-resolution biological SEM images invariably have poor SNR.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.