Abstract
This study evaluated the use of a high frequency A-mode ultrasound scanner (CUTECH DM70) for monitoring the digital artery vasospasm of Raynaud's Phenomenon (RP). The technique was evaluated on 12 RP patients and 12 age and sex matched controls. The diameter of a single digital artery was measured in these subjects over a range of finger temperatures between 14-35 degrees C. The reproducibility of the measurements was assessed by repeating the procedure in two subjects (a control and an RP patient) on 5 separate days. The results confirmed that the technique is reproducible and monitored vasospasm occurring in all 12 RP patients. In the temperature range tested, none of the controls developed digital artery vasospasm. At finger temperatures greater than 25 degrees C the response to thermal stimuli of the digital arteries in RP patients and controls was similar. However, at temperatures less than 25 degrees C a marked vasoconstriction leading to complete digital artery closure was seen in the RP patients and not in the controls. This technique may prove useful in monitoring RP patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.