Abstract

In breast MRI with diffusion-weighted imaging (DWI), fat suppression is essential for eliminating the dominant lipid signal. This investigation evaluates a combined water-excitation-spectral-fatsat method (WEXfs) versus standard spectral attenuated inversion recovery (SPAIR) in high-resolution 3-Tesla breast MRI. Multiparametric breast MRI with 2 echo-planar DWI sequences was performed in 83 patients (50.1 ± 12.6 years) employing either WEXfs or SPAIR for fat signal suppression. Three radiologists assessed overall DWI quality and delineability of 88 focal lesions (28 malignant, 60 benign) on images with b values of 800 and 1600 s/mm2, as well as apparent diffusion coefficient (ADC) maps. For each fat suppression method and b value, the longest lesion diameter was determined in addition to measuring the signal intensity in DWI and ADC value in standardized regions of interest. Regardless of b values, image quality (all p < 0.001) and lesion delineability (all p ≤ 0.003) with WEXfs-DWI were deemed superior compared to SPAIR-DWI in benign and malignant lesions. Irrespective of lesion characterization, WEXfs-DWI provided superior signal-to-noise, contrast-to-noise and signal-intensity ratios with 1600 s/mm2 (all p ≤ 0.05). The lesion size difference between contrast-enhanced T1 subtraction images and DWI was smaller for WEXfs compared to SPAIR fat suppression (all p ≤ 0.007). The mean ADC value in malignant lesions was lower for WEXfs-DWI (p < 0.001), while no significant ADC difference was ascertained between both techniques in benign lesions (p=0.947). WEXfs-DWI provides better subjective and objective image quality than standard SPAIR-DWI, resulting in a more accurate estimation of benign and malignant lesion size.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call