Abstract

DWI of the head and neck can reveal valuable information, but the effects of fat suppression and multishot acquisition on image quality have not been thoroughly investigated. We aimed to comprehensively compare the quality of head and neck DWI at 3T using 2 fat-suppression techniques, STIR, and spectral presaturation with inversion recovery, which were used with both single- and multishot EPI. Sixty-five study participants underwent 3 DWI sequences of single-shot EPI-STIR, single-shot EPI-spectral presaturation with inversion recovery, and multishot EPI-spectral presaturation with inversion recovery of the head and neck. In multiple anatomic regions, 2 independent readers assessed 5-point visual scores for fat-suppression uniformity and image distortion, and 1 reader measured the contrast-to-noise ratio and ADC. The mean visual score for fat-suppression uniformity was higher in single-shot EPI-STIR than in other sequences (all regions except for the orbital region, P < .05). The mean visual score for image distortion was higher in multishot EPI-spectral presaturation with inversion recovery than in single-shot EPI sequences (all regions, P < .001). Contrast-to-noise ratio was mostly lower in single-shot EPI-STIR than in other sequences (P < .001), and ADC was significantly higher in multishot EPI-spectral presaturation with inversion recovery than in single-shot EPI sequences (P ≤ .001). Overall, multishot EPI-spectral presaturation with inversion recovery provided the best image quality, with relatively homogeneous fat suppression, less image distortion than single-shot EPI sequences, and higher contrast-to-noise ratio than single-shot EPI-STIR. The measured ADC values can be higher in multishot EPI-spectral presaturation with inversion recovery, which necessitates cautious application of the previously reported ADC values to clinical settings.

Highlights

  • BACKGROUND AND PURPOSEDWI of the head and neck can reveal valuable information, but the effects of fat suppression and multishot acquisition on image quality have not been thoroughly investigated

  • Contrast-to-noise ratio was mostly lower in single-shot EPI–STIR than in other sequences (P Ͻ .001), and ADC was significantly higher in multishot EPI–spectral presaturation with inversion recovery than in single-shot EPI sequences (P Յ .001)

  • The measured ADC values can be higher in multishot EPI–spectral presaturation with inversion recovery, which necessitates cautious application of the previously reported ADC values to clinical settings

Read more

Summary

Objectives

We aimed to comprehensively compare the quality of head and neck DWI at 3T using 2 fat-suppression techniques, STIR, and spectral presaturation with inversion recovery, which were used with both single- and multishot EPI. The aim of this study was to perform a comprehensive comparison of the quality of DWI in the head and neck using different fat-suppression techniques and to determine the utility of the msEPI sequence. The purpose of our study was to compare both qualitative and quantitative measurements of the image qualities of the above 3 sequences

Methods
Results
Discussion
Conclusion