Abstract

To investigate the feasibility of diffusion tensor brain imaging at 0.55T with comparisons against 3T. Diffusion tensor imaging data with 2 mm isotropic resolution was acquired on a cohort of five healthy subjects using both 0.55T and 3T scanners. The signal-to-noise ratio (SNR) of the 0.55T data was improved using a previous SNR-enhancing joint reconstruction method that jointly reconstructs the entire set of diffusion weighted images from k-space using shared-edge constraints. Quantitative diffusion tensor parameters were estimated and compared across field strengths. We also performed a test-retest assessment of repeatability at each field strength. After applying SNR-enhancing joint reconstruction, the diffusion tensor parameters obtained from 0.55T data were strongly correlated ( ) with those obtained from 3T data. Test-retest analysis showed that SNR-enhancing reconstruction improved the repeatability of the 0.55T diffusion tensor parameters. High-resolution in vivo diffusion MRI of the human brain is feasible at 0.55T when appropriate noise-mitigation strategies are applied.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call