Abstract
We have recently proposed a regularized least square criterion for adaptive regularization of single photon emission computed tomography (SPECT) reconstruction with nonuniform attenuation correction. In the present study, we show that this regularization is closely related to a diffusion scheme used for Gaussian filtering. For a given value of the regularization parameter, the amount of smoothing is independent from the patient attenuation map, and it is mathematically related to the full-width at half-maximum (FWHM) of a Gaussian filter. A second regularized least square criterion is then derived for which regularization also behaves as a diffusion scheme. The new penalty is then shown to be also applicable to the weighted least square criterion, and to the Poisson maximum-likelihood criterion for positron emission tomography (PET) data (i.e., without attenuation) solved by the expectation maximization (EM) algorithm. For all these criteria, the regularization level can thus be set as the FWHM of a Gaussian filter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.