Abstract
Detailed plots of diffusion coefficients, Dm, versus % methanol in CO2 and pressure, P, are presented. Using sub-2 µm particle columns with small ID connector tubing make for large pressure drops, ∆P, across both the system and the column. This creates low average Dm, and large gradients in Dm, slowing the chromatography. The ∆P with 75 µm ID tubing varies non-linearly with flow creating U shaped plots of reduced plate height versus flow rate. Plumbing does but should never change h versus F plots. Using larger ID tubes to estimate the extra-column variance $$\sigma_{e - c}^{2}$$ of smaller tubes, by keeping the aspect ratio (L/r) constant, assumes the same constant Dm in both tubes, which is not true in SFC. Contrary to prediction, calculated tube plate count, N, indicates connector tubing is more dispersive in SFC compared to high performance liquid chromatography (HPLC). Current supercritical fluid chromatographs (SFC) have up to 20 × higher variance than appropriate for use with 2.1 and 3 mm columns ≤ 100 mm long, packed with sub-2 µm particles. Lower volume UV detector flow cells are needed for use with sub-2 µm particles. The modular nature of instruments makes it difficult to further minimize tube lengths while retaining column temperature control. Very small ID tubes, like 75 µm should probably be avoided. It remains difficult to characterize and minimize $$\sigma_{e - c}^{2}$$ .
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.