Abstract
Antibiotic-free human serum was spiked with known concentrations of liposomal amikacin and assayed on the Abbott TDx System, using polarization fluoroimmuno assay (PFIA) kits from Abbott Laboratories, Oxis and Sigma. Although all three kits gave a linear response, the Abbott and Oxis kits showed very low recovery (<21%) with only the Sigma kit giving near 100% recovery. Heating samples at 56 degrees C for 30 min improved recovery with the Abbott and Oxis kits (75-80% of target value), but decreased recovery with the Sigma kit (85% of target value). The loss of amikacin from liposomal amikacin, as measured using the Sigma kit, was related to both temperature and duration of heating, reaching a maximal loss of 21% after 1 h at 60 degrees C.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.