Abstract

Implantable devices capable of several modes of therapy will require differentiation of various ventricular tachyarrhythmias. Three methods of arrhythmia analysis, magnitude-squared coherence, ventricular rate, and irregularity of cycle length were performed for 45 episodes of induced ventricular tachyarrhythmia in 15 patients. Differentiation of monomorphic ventricular tachycardia from polymorphic ventricular tachycardia and ventricular fibrillation was possible by mean magnitude-squared coherence, less possible by rate, and not possible by beat-to-beat irregularity. Faster monomorphic ventricular tachycardia overlapped with rates of polymorphic ventricular tachycardia and ventricular fibrillation. Differentiation of polymorphic ventricular tachycardia and ventricular fibrillation was not possible by rate or irregularity. A progressive decrease in mean magnitude-squared coherence from monomorphic ventricular tachycardia to polymorphic ventricular tachycardia to ventricular fibrillation strengthens previous observations that coherence is a measure of rhythm "organization."

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.