Abstract

m-Cresol is listed as a priority controlled contaminant in many countries, but it is very difficult to accurately determine isomeric cresols due to their incomplete chromatographic separation on commercially available chromatographic columns and their nearly identical mass spectra. Silylation of isomeric cresols was carried out by treatment with N-methyl-N-(trimethylsilyl)trifluoroacetamide. The formed trimethyl(tolyloxy)silanes were analyzed by gas chromatography/mass spectrometry (GC/MS). Theoretical calculations were carried out with the Gaussian 03 program using the density functional theory (DFT) method at the B3LYP/6-311 + G(2d,p) level. The derivatives of three isomeric cresols and six isomeric xylenols have been completely separated on an HP-5MS capillary column within a GC run of only 10 minutes. In addition, the derivative o-cresol can be very easily differentiated from its isomers due to its characteristic base peak ion at m/z 91 in electron ionization (EI)-MS. DFT calculation results indicated that the formation of the abundant fragment ion at m/z 91 is attributed to a facile dissociation pathway involving the shift of a neighboring phenylmethyl hydrogen atom in EI-MS of trimethyl(o-tolyloxy)silane. Silylation provides a promising solution for simultaneous determination of isomeric cresols and isomeric xylenols.