Abstract

Dynamic Contrast Enhanced Magnetic Resonance Imaging (DCE-MRI) is an effective diagnostic modality for symptomatic breast disease. However, its role in evaluating clinically occult disease associated with mammographically detected microcalcification remains unclear. Women recalled following screening mammography with microcalcification had DCE-MRI examination of the breast. The data were evaluated subjectively and objectively using both empirical and 2-compartment pharmacokinetic modelling techniques to evaluate signal intensity parameters. Eighty-eight patients aged 50-75 years (median 58) were recruited. Comparing malignant and benign lesions, the mean values in arbitrary units for the enhancement index at 1 min in the most enhancing 9-pixel square +/-1 standard deviation were 0.61+/-0.40 vs. 0.22+/-0.26 p=<0.001 with sensitivity, specificity, PPV, NPV and accuracy of 80.0%, 82.4%, 57.1%, 93.3% and 81.8%, respectively. The corresponding values attained by the radiologist were 75.0%, 89.7%, 68.2%, 92.4% and 86.4%. DCE-MRI is able to differentiate malignant from benign clinically occult lesions associated with microcalcification and may therefore offer an alternative to open surgical biopsy for women with equivocal findings following initial triple assessment for microcalcification in the breast.