Abstract
PurposeTo differentiate between keratoconus and contact lens-related corneal warpage by combining focal change patterns in anterior corneal topography, pachymetry, and epithelial thickness maps.MethodsPachymetry and epithelial thickness maps of normal, keratoconus, and warpage, and forme fruste keratoconus (FFK) eyes were obtained from a Fourier-domain optical coherence tomography (OCT). Epithelial pattern standard deviation (PSD) was calculated and combined with two novel indices, the Warpage Index and the Anterior Ectasia Index, to differentiate between normal, keratoconus, and warpage eyes. The values of the three parameters were compared between groups.ResultsThe study included 22 normal, 31 keratoconic, 11 warpage, and 8 FFK eyes. The epithelial PSD was normal (< 0.041) for 100% normal eyes and abnormal (> 0.041) for 100% of keratoconic eyes, 81.8% of warpage eyes, and 87.5% of FFK eyes. The Anterior Ectasia Index of normal eyes (1.66 ± 0.74) was significantly lower than that for the keratoconus eyes (17.5 ± 7.17), the warpage eyes (2.98 ± 1.69), and the FFK eyes (6.95 ± 5.86). The Warpage Index was positive in all warpage eyes and negative for all keratoconic and FFK eyes except three wearing rigid gas-permeable contact lens.ConclusionsThe epithelial PSD can distinguish normal from keratoconus or warpage, but does not distinguish between these two conditions. The Anterior Ectasia Index is abnormal in keratoconus but not warpage. The Warpage Index is positive for warpage and negative for keratoconus, except in cases where keratoconus and warpage coexist. Together, the three parameters are strong tripartite discriminators of normal, keratoconus, and warpage.
Highlights
ObjectivesThe purpose of this study is to differentiate keratoconus from contact lens-related warpage by combining focal change patterns of several corneal maps: anterior topography, pachymetry, and epithelial thickness
The Warpage Index was positive in all warpage eyes and negative for all keratoconic and forme fruste keratoconus (FFK) eyes except three wearing rigid gas-permeable contact lens
The epithelial pattern standard deviation (PSD) can distinguish normal from keratoconus or warpage, but does not distinguish between these two conditions
Summary
The purpose of this study is to differentiate keratoconus from contact lens-related warpage by combining focal change patterns of several corneal maps: anterior topography, pachymetry, and epithelial thickness
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.