Abstract

To determine the variation of Af along the axial length of rotary nickel-titanium endodontic files (RNEF). Three commercial brands of 4% taper RNEF: GTX (#20, 25 mm, Dentsply Tulsa Dental Specialties, Tulsa, OK, USA), K3 (#25, 25 mm) and TF (Twisted File #25, 27 mm) (Sybron Kerr, Orange, CA, USA) were cut into segments at 4 mm increment from the working tip. Regional specimens were measured for differential heat-flow over thermal cycling, generally with continuous heating or cooling (5 °C/min) and 5 min hold at set temperatures (start, finish temperatures): GTX: −55, 90 °C; K3: −55, 45 °C; TF: −55, 60 °C; using differential scanning calorimeter. This experiment demonstrated regional differences in Af along the axial length of GTX and K3 files. Similar variation was not obvious in the TF samples. A contributory effect of regional difference in strain-hardening due to grinding and machining during manufacturing is proposed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.