Abstract

A 10 V programmable Josephson voltage standard has enabled sine waves with voltages up to 7 V RMS to be accurately measured with a differential sampling measurement technique. Expanding the voltage range for this technique enables the direct calibration of the low-frequency ranges of commercial calibrators in the ac voltage mode. This paper reviews the practically achievable performance and challenges of the differential sampling measurement technique that arise when measuring RMS voltages greater than a few volts. A relative Type A uncertainty of 4 parts in $10^{7}$ was achieved with the technique when measuring a 7 V RMS sine wave generated by a calibrator at 62.5 Hz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.