Abstract

Expressions for reversible differential pulse (DP) voltammograms at microdisk electrodes were established from experimental results as a function of pulse electrolysis parameters. Three expressions were computed numerically, depending on the electrode radius and varying the pulse interval, τ. They were compared to theoretical and numerical expressions established previously in literature reports for macrodisk and microcylinder electrodes. The other aim of this work was to determine the minimum value of the pulse interval for which DP voltammograms observed at microdisk electrodes are not distorted. It appeared that no distortion is observed even at very short values of τ, meaning that microdisk electrodes are a good alternative to macroelectrodes in order to reduce time analysis in DPV. DP measurements were carried out at platinum electrodes for two reversible reactions: hexacyanoferrate (II) and iodide oxidation reaction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.