Abstract

The selenium content in blood was determined using the hydrogen catalytic peak. This peak at −1.1 V was obtained in the presence of selenium and molybdenum at pH values of 1–4 in different buffers. For the determination of selenium, the Mo(VI) concentration has to be ∼100–200 times higher than the selenium present. The linear domain range of selenium is 1×10 −6–5×10 −9 M. The interference of zinc is eliminated by the addition of EDTA at pH 3.5 acetate buffer. The method was applied to 1.0 ml of digested blood, and 620±44 μg l −1 Se and 7.15 mg l −1 Zn could be determined with a 90% ( n=6) confidence interval.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.